A reliable non-separation fluorescence quenching assay for total glycated serum protein: a simple alternative to nitroblue tetrazolium reduction:

2000 
A simple non-separation assay for the measurement of total glycated serum protein is described. It was found that the fluorescence intensity of a solution of a fluorescein-boronic acid derivative was quenched in proportion to the amount of serum added. This led to the development of an assay in which 10 µL of serum is added to 4 mL of a solution of the fluorescein-boronic acid derivative and the fluorescence intensity is measured after 15 min. The results, as measured by drop in fluorescence intensity, calibrated by a single standard, were compared with the results for nitroblue tetrazolium (NBT) reduction of fructosamine and showed good correlation (r=0·936, n=114). The intra-assay precision (seven samples each measured 10 times) was less than 2·1% (concentration range 190-660 µmol/L); inter-assay precision for seven samples in 10 assays was less than 2·5% (over the same concentration range). Dilution of serum that had a high concentration of total glycated protein showed the assay to be linear. Serum sa...