Evaluating noise in digitized radiographic images by means of histogram
2006
To evaluate the performance of three digital devices regarding the noise added to digital radiographic images containing different optical densities. METHODS: A radiographic image was digitized repeatedly ten times using two scanners (HP 4c/T and HP 5370C) and a digital camera (Nikon 990). A histogram tool measured a mean pixel value and the standard deviation of the region of interest in each image. Both values were used to calculate the image noise at the different optical densities. RESULTS: The noise values found were different for all devices and optical densities. There was a statistically significant difference (p OBJETIVOS: Avaliar tres equipamentos digitais em relacao ao ruido agregado as imagens radiograficas digitalizadas contendo diferentes densidades opticas. MATERIAL AND METODO: Uma imagem radiografica foi digitalizada sequencialmente dez vezes usando dois escaneres (HP 4c/T and HP 5370C) e uma câmera digital (Nikon 990). Por meio do histograma foram medidos os valores de pixels e os desvios-padroes da regiao de interesse de cada imagem. Ambos valores foram utilizados para o calculo do ruido nas diferentes densidades opticas. RESULTADOS: Os valores encontrados para o ruido foram diferentes para cada equipamento e para cada densidade optica. Houve uma diferenca estatistica significante entre os valores de ruido encontrados para o escaner HP 4c/T e a câmera digital (p
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
15
References
4
Citations
NaN
KQI