Quality Improvement of Environmental Secondary Electron Detector Signal Using Helium Gas in Variable Pressure Scanning Electron Microscopy

2007 
The quality of the image signal obtained from the environmental secondary electron detector (ESED) employed in a variable pressure (VP) SEM can be dramatically improved by using helium gas. The signal-to-noise ratio (SNR) increases gradually in the range of the pressures that can be used in our modified SEM. This method is especially useful in low-voltage VP SEM as well as in a variety of SEM operating conditions, because helium gas can more or less maintain the amount of unscattered primary electrons. In order to measure the SNR precisely, a digital scan generator system for obtaining two images with identical views is employed as a precondition. SCANNING 29: 225-229, 2007. © 2007 Wiley Periodicals, Inc.