Direct analysis – no sample preparation – of bioavailable cortisol in human plasma by weak affinity chromatography (WAC)

2017 
Abstract Pre-analytical treatment of blood plasma is a time consuming and often rate limiting step in the workflow of LC/MS analysis. We present in this pilot study a new approach for quantitative LC/MS based on weak affinity chromatography (WAC) of crude plasma. The steroid hormone cortisol was selected as a clinically relevant biomarker, as it currently requires extensive pre-analytical preparation. A WAC unit with saturating, immobilized albumin as a prototypic weak binder was used in combination with an ion-funnel MS/MS detector to perform zonal affinity chromatography of cortisol directly from a plasma sample, followed by quantitative multiple reaction monitoring (MRM). This procedure also allowed us to determine the amount of bioavailable cortisol in the clinical plasma sample which is of significant therapeutic interest. This WAC-MS approach showed an excellent correlation (R 2  = 0.86 (P