Evaluation of the Xpert HCV Viral Load Finger-Stick Point-of-Care Assay

2018 
: Point-of-care hepatitis C virus (HCV) RNA testing is advantageous, enabling diagnosis of active infection in a single visit. This study evaluated the sensitivity and specificity of the Xpert HCV Viral Load Finger-Stick assay (Xpert HCV VL FS) for HCV RNA detection (finger-stick) and the Xpert HCV Viral Load assay (plasma) compared with the Abbott RealTime HCV Viral Load assay by venepuncture. Plasma and finger-stick capillary whole-blood samples were collected from participants in an observational cohort in Australia. Of 223 participants enrolled, HCV RNA was detected in 40% of participants (85 of 210) with available Xpert HCV Viral Load testing. Participants receiving HCV therapy were excluded from subsequent analyses (n = 16). Sensitivity of the Xpert HCV Viral Load assay for HCV RNA quantification in plasma collected by venepuncture was 100.0% (95% confidence interval [CI] 96.9%-100.0%) and specificity was 100.0% (95% CI, 94.4%-100.0%). Sensitivity of the Xpert HCV VL FS assay for HCV RNA quantification in samples collected by finger-stick was 100.0% (95% CI, 93.9%-100.0%) and specificity was 100.0% (95% CI, 96.6%-100.0%). The Xpert HCV VL FS test can accurately detect active infection from a finger-stick sample in 1 hour allowing single-visit HCV diagnosis.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    32
    References
    58
    Citations
    NaN
    KQI
    []