Congruency of scapula locking plates: implications for implant design.

2012 
: We conducted a study to evaluate the congruency of fit of current scapular plate designs. Three-dimensional image-processing and -analysis software, and computed tomography scans of 12 cadaveric scapulae were used to generate 3 measurements: mean distance from plate to bone, maximum distance, and percentage of plate surface within 2 mm of bone. These measurements were used to quantify congruency. The scapular spine plate had the most congruent fit in all 3 measured variables. The lateral border and glenoid plates performed statistically as well as the scapular spine plate in at least 1 of the measured variables. The medial border plate had the least optimal measurements in all 3 variables. With locking-plate technology used in a wide variety of anatomical locations, the locking scapula plate system can allow for a fixed-angle construct in this region. Our study results showed that the scapular spine, glenoid, and lateral border plates are adequate in terms of congruency. However, design improvements may be necessary for the medial border plate. In addition, we describe a novel method for quantifying hardware congruency, a method that can be applied to any anatomical location.