Analytical validation of a second-generation immunoassay for the quantification of N-terminal pro–B-type natriuretic peptide in canine blood

2015 
N-terminal pro–B-type natriuretic peptide (NT-proBNP) has been shown to have clinical utility as a biomarker in dogs with heart disease. There were several limitations associated with early diagnostic assay formats including a limited dynamic range and the need for protease inhibitors to maintain sample stability. A second-generation Cardiopet® proBNP enzyme-linked immunosorbent assay (IDEXX Laboratories Inc., Westbrook, Maine) was developed to address these limitations, and the present study reports the results of the analytical method validation for the second-generation assay. Coefficients of variation for intra-assay, interassay, and total precision based on 8 samples ranged from 3.9% to 8.9%, 2.0% to 5.0%, and 5.5% to 10.6%, respectively. Analytical sensitivity was established at 102 pmol/l. Accuracy averaged 102.0% based on the serial dilutions of 5 high-dose canine samples. Bilirubin, lipids, and hemoglobin had no effect on results. Reproducibility across 3 unique assay lots was excellent with an a...