Recovery of CVD diamond detectors using laser double pulses

2006 
A 5 x 0.25mm Chemical Vapor Deposited (CVD) diamond detector, with a voltage bias of +250 V, was excited by a 400 nm laser (3.1 eV photons) in order to study the saturation of the wafer and its associated electronics. In a first experiment, the laser beam energy was increased from a few tens of a pJ to about 100 μJ, and the signal from the diamond was recorded until full saturation of the detection system was achieved. Clear saturation of the detection system was observed at about 40V, which corresponds with the expected saturation at 10% of the applied bias (250V). The results indicate that the interaction mechanism of the 3.1 eV photons in the diamond (E bandgap = 5.45 eV) is not a multi-photon process but is linked to the impurities and defects of the crystal. In a second experiment, the detector was irradiated by a saturating first laser pulse and then by a delayed laser pulse of equal or smaller amplitude with delays of 5, 10, and 20ns. The results suggest that the diamond and associated electronics recover within 10 to 20ns after a strong saturating pulse.