Catheter-based Arterial Input Function Determination for Myocardial Perfusion Measurements.

2020 
Abstract The quantification of myocardial perfusion with contrast agent (CA) tracers requires the precise knowledge of the arterial input function (AIF). In this study a method for MR-guided vascular interventions is evaluated that determines the AIF via an active tracking catheter during targeted CA injection. A phantom experiment with a dialysis filter was conducted to measure the AIF using an active catheter and a dynamic image series as reference. To compensate for dilution and coil sensitivity effects, correction methods were developed for the catheter-based AIF determination. From the dynamic MR measurements in the perfusion phantom quantitative perfusion maps were calculated by a deconvolution of the measured CA concentration with the AIF, and additional flow measurements were used to normalize the perfusion map. The signal-time-curves of the measured AIF using the catheter-based and imaging-based methods agree while the absolute values differ by a scaling factor of about 9. After normalization to the surrounding flow, both perfusion techniques are in excellent agreement. Catheter-based AIF measurements are feasible but require an additional normalization which can be determined from a flow measurement. The technique might enable faster perfusion measurements during cardiovascular interventions.