Evaluation of interference factors in creatinine measurement by using new guideline CLSI EP7-A3

2020 
Objective To investigate the interfering factors in the determination of creatinine(Cr) using the American Clinical Laboratory Standards Association (CLSI) EP7-A3 document. Methods According to the CLSI EP7-A3 document, fresh serum (no hemolysis, lipemia, and jaundice) was used on the day of the experiment and confirmed the interfering substances through the pairing difference experiment and the point-to-point analysis method was used in the dose effect experiment to clarify the difference of interfering substances. Results Triglyceride (16.94 mmol/L), dobutamine hydrochloride (4.01 μmol/L), ascorbic acid (298 μmol/L) did not interfere with the determination of Cr. Free bilirubin (684 μmol/L), conjugated bilirubin (684 μmol/L), calcium hydroxybenzene sulfonate (144 μmol/L) and hemoglobin (10 g/L) were used as the maximum concentrations of interferences for the dose effect test, the results showed that the above interferences had negative interference on the determination of Cr. Conclusion According to EP7-A3, it is valuable to evaluate the interference factors of creatinine determination. Key words: Interference experiment; Creatinine