Development and validation of an automated algorithm to evaluate the abundance of bubbles in small bowel capsule endoscopy
2018
Background and study aims Bubbles can impair visualization of the small bowel (SB) mucosa during capsule endoscopy (CE). We aimed to develop and validate a computed algorithm that would allow evaluation of the abundance of bubbles in SB-CE still frames. Patients and methods Two sets of 200 SB-CE normal still frames were created. Two experienced SB-CE readers analyzed both sets of images twice, in a random order. Each still frame was categorized as presenting with Results Both SURF and GLCM algorithms had high operating points (Se and Sp over 90 %) and a perfect reproducibility (κ = 1). The validation step showed the GLCM detector strategy had the best diagnostic performances, with a Se of 95.79 %, a Sp of 95.19 %, and a calculation time of 0.037 seconds per frame. Conclusion A computed algorithm based on a GLCM detector strategy had high diagnostic performance allowing assessment of the abundance of bubbles in SB-CE still frames. This algorithm could be of interest for clinical use (quality reporting) and for research purposes (objective comparison tool of different preparations).