Quantra™ should be considered a tool for two-grade scale mammographic breast density classification

2016 
Objective:To assess the agreement between Quantra™ (Hologic Inc., Bedford, MA) and Breast Imaging Reporting and Data Systems (BI-RADS®) and the performance of Quantra at reproducing BI-RADS mammographic breast density (MBD) assessment.Methods:MBD assessment was performed using Quantra and BI-RADS. BI-RADS assessment was performed in two phases (1314 and 292 cases, respectively). Kappa was used to assess the interreader agreement and the agreement between Quantra and BI-RADS, and receiver-operating characteristics analysis was used to assess the performance of Quantra at reproducing BI-RADS rating.Results:Agreement (weighted kappa) between BI-RADS and Quantra in Phase 1 was 0.75 [95% confidence interval (CI): 0.73–0.78] and 0.85 (95% CI: 0.80–0.90) on four- and two-grade scales, respectively. The corresponding agreement in Phase 2 was 0.79 (95% CI: 0.75–0.84) and 0.84 (95% CI: 0.79–0.87) using the majority report. In Phase 1, Quantra demonstrated 93.2% sensitivity and 86.1% specificity for BI-RADS on a two...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    38
    References
    8
    Citations
    NaN
    KQI
    []