Bilateral comparison of 10 V standards between the NSAI?NML (Ireland) and the BIPM, March to April 2011 (part of the ongoing BIPM key comparison BIPM.EM-K11.b)

2011 
As a part of the ongoing BIPM key comparison BIPM.EM-K11.b, a comparison of the 10?V voltage reference standards of the BIPM and the National Standards Authority of Ireland?National Metrology Laboratory (NSAI?NML), Dublin, Ireland, was carried out from March to April 2011. Two BIPM Zener diode-based travelling standards (Fluke 732B) were transported by freight to NSAI?NML. At NSAI?NML, the 10?V output EMF of each travelling standard was measured by direct comparison with a group of characterized Zener diode-based electronic voltage standards. At the BIPM, the travelling standards were calibrated before and after the measurements at NSAI?NML, with the Josephson Voltage Standard. Results of all measurements were corrected for the dependence of the output voltages on internal temperature and ambient pressure. The comparison results show that the voltage standards maintained by NSAI?NML and the BIPM were equivalent, within their stated expanded uncertainties, on the mean date of the comparison. Main text. To reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/. The final report has been peer-reviewed and approved for publication by the CCEM, according to the provisions of the CIPM Mutual Recognition Arrangement (MRA).