Improved coronary artery contrast enhancement using noise-optimised virtual monoenergetic imaging from dual-source dual-energy computed tomography

2019 
Abstract Purpose To define optimal kiloelectron volt (keV) settings for virtual monoenergetic imaging (VMI) reconstruction at dual-energy coronary computed tomography angiography (DE-CCTA). Method Fifty-one DE-CCTA data sets (33 men; mean age, 63.9 ± 13.2 years) were reconstructed as standard linearly-blended images (F_0.6; 60% of 90 kVp, 40% of 150 kVpSn), and with traditional (VMI) and noise-optimised (VMI+) algorithms from 40 to 100 keV in 10-keV intervals. Objective image quality was assessed with signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) measurements. Three observers subjectively evaluated vascular contrast, image sharpness, noise and delineation of coronary plaques. Results Median values for objective image analysis were highest in VMI + series at 40 keV (SNR, 44.5; CNR: 33.5), significantly superior (allp  Conclusions DE-CCTA image reconstruction with 40-keV VMI + allows for significant improvement of both objective and subjective image quality.