The influence of different technique factors on image quality of lumbar spine radiographs as evaluated by established CEC image criteria

2000 
In this study we have investigated the image quality of lumbar spine radiographs taken after recording technical and physical parameters. Two technical parameters were altered, tube voltage (70 kV and 90 kV for the anteroposterior (AP) projection and 77 kV and 95 kV for the lateral projection) and sensitivity of the film-screen system (sensitivity class 400 and 600). In total, 85 images were included in the study. Entrance surface dose (ESD) was measured using thermoluminescent dosemeters. The mean value of ESD for the different technique groups varied between 1.9 mGy (90 kV, sensitivity class 400) and 4.6 mGy (70 kV, sensitivity class 400) for the AP projection, and between 6.4 mGy (95 kV, sensitivity class 600) and 20.4 mGy (70 kV, sensitivity class 400) for the lateral projection. Image criteria given in the "European Guidelines on Quality Criteria for Radiographic Images" were used to assess image quality. Two evaluation methods have been employed. A straightforward scoring of fulfilled image criteria...