Electrochemical behavior of two austenitic stainless steel biomaterials Comportamento eletroquímico de dois aços inoxidáveis austeníticos utilizados como biomateriais

2010 
Potentiodynamic and potentiostatic polarization measurements were conducted in 0.9% NaCl solution to investigate the localized corrosion susceptibility of two austenitic stainless steel biomaterials: a low nitrogen, according to ASTM F 138 - the metallic material most widely utilized today in orthopedic applications; and a nitrogen- and niobium-bearing stainless steel, according to ISO 5832-9, which has shown a promising potential as a substitute of the F 138 steel for more severe loading applications and longer times inside the human body. The polarization tests revealed that the ISO 5832-9 steel is more corrosion resistant than the F 138. The critical pitting potential of the ISO 5832-9 steel could not be observed in the cyclic polarization curve up to the value of potential corresponding to its transpassivity. The potentiostatic scratch test confi rmed the superiority of the ISO 5832-9 steel, which heal the mechanically damaged passive fi lm at applied potential as high as 800 mV (SCE). Scanning electron microscopy (SEM) examination confi rmed the presence of corrosion pits - lacelike pits - in a stable growth stage on the surface of F 138 steel and the absence of such pits in the specimens of the ISO 5832-9