Reduction of procedure duration and radiation exposure with a dedicated inner lumen mapping catheter during pulmonary vein cryoablation.

2013 
Background The Achieve catheter (AC; Medtronic-CryoCath, Pointe-Claire, Canada) is a circular mapping catheter introduced through the lumen of the cryoballoon (CB) catheter which is safe and effective to both navigate the CB to the pulmonary veins (PV) and allow PV potential recording during PV cryoablation. The aim of this study was to evaluate the impact of the use of the AC on procedural outcomes. Methods Sixteen consecutive patients (14 men) underwent AC-guided PV isolation (PVI) for drug-refractory paroxysmal atrial fibrillation (AF; AC group). Clinical and procedural data of these patients were compared to those obtained from 16 consecutive patients who had undergone PVI for paroxysmal AF with the regular “single transseptal” approach (control group). Results Clinical characteristics of patients enrolled in both groups did not differ significantly. In the AC group, 64 PVs were targeted using a single 28-mm (n = 13) or 23-mm (n = 3) CB catheter with PVI achieved in 62 PVs (97%). In the control group, 66 PVs were targeted using a single 28-mm (n = 12) or 23-mm (n = 4) CB catheter with PVI achieved in 62 PVs (94%), (P = non-significant for CB size and PVI rate). Procedure duration and fluoroscopy time were significantly reduced in AC group compared to control group (96.6 ± 26 minutes vs 125.9 ± 25 minutes, P = 0.003 and 24.4 ± 10 minutes vs 32.6 ± 11 minutes, P = 0.04, respectively). Conclusion The use of the AC significantly reduced procedure duration and radiation exposure during PVI with the CB technique. (PACE 2013; 36:24–30)
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    16
    References
    11
    Citations
    NaN
    KQI
    []