Rapid quantification of lisinopril in human plasma by liquid chromatography/tandem mass spectrometry
2007
An assay based on protein precipitation and liquid chromatography/tandem mass spectrometry (LC-MS/MS) has been developed and validated for the quantitative analysis of lisinopril in human plasma. After the addition of enalaprilat as internal standard (IS), plasma samples were prepared by one-step protein precipitation using perchloric acid followed by an isocratic elution with 10 mm ammonium acetate buffer (pH adjusted to 5.0 with acetic acid)–methanol (70:30, v/v) on a Phenomenex Luna 5µC18 (2) column. Detection was performed on a triple-quadrupole mass spectrometer utilizing an electrospray ionization (ESI) interface operating in positive ion and selected reaction monitoring (SRM) mode with the precursor to product ion transitions m/z 406246 for lisinopril and m/z 349206 for enalaprilat. Calibration curves of lisinopril in human plasma were linear (r = 0.9973−0.9998) over the concentration range 2–200 ng/mL with acceptable accuracy and precision. The limit of detection and lower limit of quantification in human plasma were 1 and 2 ng/mL, respectively. The validated LC-MS/MS method has been successfully applied to a preliminary pharmacokinetic study of lisinopril in Chinese healthy male volunteers. Copyright © 2007 John Wiley & Sons, Ltd.