Misassembled bellows housing leading to failure of mechanical ventilation during general anesthesia.

2012 
Demographics Case Title: Circuit Leak (Ventilator) Patient Name: Kaman Cipi Scenario Name: Misassembled Bellows Housing Simulation Developers: Nahel N. Saied, MB, BCh, Matthew B. Weinger, MD, and Raymond Booker Date of Development: December 20, 2010 Appropriate for the following learning groups: Faculty: CME Residents: PGY 2, 3, 4 Specialties: Anesthesiology Nurse Anesthesia Faculty: CEU Nursing Students (yr): 2 Other: Anesthesia Machine Technician/Representative Note: Depending on the target student learner, attention to patient care will be adjusted: 1. If the learner is the anesthesia provider: & The beginner learner will be given the opportunity to delegate patient care to a colleague (confederate), so he/she can trouble shoot the machine issue. & The experienced learner could assist the technician (confederate) in trouble shooting the problem in addition to providing patient care (no help available to assume patient care, see Discussion section). The confederate will appear lost and unable to make a diagnosis or formulate alternative plans.