Integrated microfluidic biochips for DNA microarray analysis

2006 
A fully integrated and self-contained microfluidic biochip device has been developed to automate the fluidic handling steps required to perform a gene expression study of the human leukemia cell line (K-562). The device consists of a DNA microarray semiconductor chip with 12,000 features and a microfluidic cartridge that consists of microfluidic pumps, mixers, valves, fluid channels and reagent storage chambers. Microarray hybridization and subsequent fluidic handling and reactions (including a number of washing and labeling steps) were performed in this fully automated and miniature device before fluorescent image scanning of the microarray chip. Electrochemical micropumps were integrated in the cartridge to provide pumping of liquid solutions. A micromixing technique based on gas bubbling generated by electrochemical micropumps was developed. Low-cost check valves were implemented in the cartridge to prevent cross-talk of the stored reagents. A single-color transcriptional analysis of K-562 cells with a...
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    39
    References
    31
    Citations
    NaN
    KQI
    []