EARLY LOOSENING OF THE ULNAR IMPLANT WITH THE COONRAD-MORREY TOTAL ELBOW PROSTHESES

2004 
Purpose: Loosening remains a problem with semi-constrained total elbow prostheses. The trend in recent years has been to improve prosthesis design to achieve stability of the humeral implant. We report a small series of nine Coonrad-Morrey total elbow prostheses where three early loosenings were observed in the ulnar implant. We attempt to analyse the causes and present a review of the recent literature.Material and methods: Nine patients, mean age 60 years, age range 57–63 years, underwent total elbow arthroplasty with a Coonrad-Morrey prosthesis for rheumatoid disease (n=5), stiff degenerative joints after trauma (n=3, flexion-extension 20°), floating joint after trauma (n=1). The posterolateral approach described by Bryan and Morrey was used for eight elbows and the posterior approach for one. Clinical and radiological results were assessed with the performance index and the Mayo clinic score respectively.Results: Mean follow-up was 3.6 years (1.5–4.7). Outcome was very good or good for seven elbows (s...