Summary A new sensitive and specific isocratic RP-HPLC-UV method was developed and validated for the determination of ondansetron in rabbit plasma using risperidone as an internal standard (IS). The sample preparation involved a simple deprotenization procedure with a mixture of 1 mL of acetonitrile and 50 μL of 10% w/υ zinc sulfate. Analysis was performed on a Phenomenex CN column (250 mm × 4.6 mm, 5 μm) with 50 mM ammonium acetate (pH 3.5) and acetonitrile (35:65, υ/υ) as mobile phase at a flow rate of 1.0 mL min−1. Column eluent was monitored at 310 nm. The calibration curve was linear over the concentration range of 25–1000 ng mL−1 (r2 = 0.9999) with a limit of quantification (LOQ) 25 ng mL−1. The intraday and interday precision and accuracy were between 0.93% and 3.41% and −3.63% and 1.01%, respectively. The mean recoveries of ondansetron and risperidone were 85.87% and 99.80%, respectively. Ondansetroncontaining plasma samples were stable at −20°C for 14 days. The validated method was successfully a...