Image quality and radiation dose in 256-slice cardiac computed tomography: Comparison of prospective versus retrospective image acquisition protocols

2011 
Abstract Purpose To assess coronary artery image quality and patient radiation exposure in patients who underwent clinically indicated 256-slice CTA. Methods Consecutive patients ( n  = 193) underwent 256-slice CTA, using (1) retrospective gating without radiation dose modulation, (2) retrospective gating with radiation dose modulation and (3) prospective gating. Image quality was determined by consensus of two experienced observers using a 5-grade scale. The effective dose was calculated. Results In all patients, CTA was performed without adverse events. Retrospective CTA was assessed in 39 patients with and 39 without dose modulation, while 115 patients underwent prospective CTA. Heart rate was related to image quality with all protocols ( r  = 0.46, p p Conclusions In patients with heart rates up to 75 bpm prospective CTA should be the first choice acquisition protocol. For heart rates >75 bpm, retrospective CTA with dose modulation should be considered.