Detailed Assessment of Low-Voltage Zones Localization by Cardiac MRI in Patients With Implantable Devices

2021 
Abstract Objectives The purpose of this study was to assess the performance and limitations of low-voltage zones (LVZ) localization by optimized late gadolinium enhancement (LGE) cardiac magnetic resonance (CMR) scar imaging in patients with cardiac implantable electronic devices (CIEDs). Background Scar evaluation by LGE-CMR can assist ventricular tachycardia (VT) ablation, but challenges with electroanatomical maps coregistration and presence of imaging artefacts from CIED limit accuracy. Methods A total of 10 patients underwent VT ablation and preprocedural LGE-CMR using wideband imaging. Scar was segmented from CMR pixel signal intensity maps using commercial software (ADAS-VT, Galgo Medical) with bespoke tools and compared with detailed electroanatomical maps (CARTO). Coregistration of EP and imaging-derived scar was performed using the aorta as a fiducial marker, and the impact of coregistration was determined by assessing intraobserver/interobserver variability and using computer simulations. Spatial smoothing was applied to assess correlation at different spatial resolutions and to reduce noise. Results Pixel signal intensity maps localized low-voltage zones (V  0.86 for area under the receiver-operating characteristic curve). Spatial smoothing improved localization of LVZ. Results for LVZ with V  Conclusions In patients with CIEDs, novel wideband CMR sequences and personalized coregistration strategies can localize LVZ with good accuracy and may assist VT ablation procedures.