Use of XR-QA2 radiochromic films for quantitative imaging of a synchrotron radiation beam

2015 
In the framework of an ongoing project, promoted by INFN, at the SYRMEP beamline of the ELETTRA synchrotron radiation facility (Trieste, Italy) for phase-contrast breast X-ray computed tomography, the assessment of the dose to the breast is one of the issues, requiring the determination of the distribution of X-ray incident photon fluence. This work investigates the use of XR-QA2 radiochromic films for quantitative imaging of the synchrotron radiation (SR) beam. XR-QA2 films were irradiated in a plane transverse to the beam axis, with a monochromatic beam of energy of 28, 35, 38 or 40 keV. The response of the radiochromic film was calibrated in terms of average air kerma measured with an ionization chamber. The net reflectance of the exposed film was then converted to photon fluence per unit air kerma (mm−2mGy−1). The SR beam profile was acquired also with a scintillator (GOS) based, fiber optic coupled CCD camera as well as with a scintillator based flat panel detector. Horizontal and vertical line profiles acquired with the radiochromic films show the 2D distribution of the beam intensity, with variations in the order of 15–20% in the horizontal direction. The response of the radiochromic film is comparable to that of the other imaging detectors, within less than 5% variation.