Irrigation eliminates smoke formation in laser laparoscopic surgery: ex vivo results.

2008 
Abstract To evaluate the desmoke effects of irrigation on surgical smoke generated during laser laparoscopic surgery using ex vivo model. A hand-piece was devised as both an irrigation tube and a laser fiber holder. The laser system used was a diode pumped continuous wave Thulium YAG laser (Lisa laser products OHG, Katlenburg-Lindau, Germany) emitting at 2.01-mum through a quartz fiber with a core diameter of 365 mum. Using a transparent container as experimental model (approximately 3 L in volume), 6 irrigation rates (0, 20, 40, 60, 80, and 100 mL/min) were tested combining with 4 laser power settings (20, 30, 40, and 50 W). To evaluate the desmoke efficiency of irrigation, lasing time with effective visibility under different combination of irrigation rates and laser powers were documented, and corresponding smoke generation rate were calculated (=3 L/lasing time). Analysis of variance was used to compare difference and significance was indicated at P