A sensitometric comparison of four dental X-ray films and their diagnostic accuracy.

1997 
OBJECTIVES To compare the sensitometric properties and accuracy in the diagnosis of approximal caries of two E-speed (Ektaspeed Plus and EV 57) and two D-speed (Ultra-speed and DV 57) films. METHODS Speed and contrast for the four films were determined following 30 exposures for each film. Base plus fog density was measured on unexposed films and the increase was studied over a 18-month period. Seventy-five extracted premolars with known pathology were radiographed under standardized conditions and the radiographs assessed by nine observers for approximal caries. Receiver Operating Characteristic (ROC) curve technique was used to analyse the diagnostic accuracy. RESULTS The sensitometric analysis showed that the Ektaspeed Plus film had the highest speed of the four films. At a density of 1.0 it was 2.28 times faster than Ultra-speed. The contrast of Ektaspeed Plus was comparable with that of the other films and base plus fog density plateaued at 0.4 ten months before expiry date. There were no statistical...