Detection of plasmin based on specific peptide substrate using acoustic transducer

2016 
Abstract In this work we report the detection of plasmin protease by means of the thickness shear mode (TSM) acoustic method. The biorecognition element consists of a peptide substrate (PS) specific to plasmin immobilized on a piezoelectric quartz crystal electrode. After enzymatic reaction with plasmin, it cleaves a short fragment of the peptide causing increase in the resonance frequency of the piezo crystal. Plasmin was detected in the range of concentrations 1–20 nM, a target interval in which its presence presumably affects the quality of milk. The PS exhibited negligible response against to similar protease trypsin. This has been confirmed also by electrochemical detection method. Limit of detection of this acoustic transducer was found to be 0.65 nM. Formation of the sensing surface and kinetic effect of plasmin on the peptide substrate was studied by atomic force microscopy (AFM). The PS response was also validated in pretreated milk samples spiked by known concentrations of plasmin achieving an average recovery of 63 ± 0.6%.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    33
    References
    13
    Citations
    NaN
    KQI
    []