Early polyethylene wear and osteolysis with ABG acetabular cups (7- to 12-year follow-up)

2006 
We reviewed 81 consecutive ABG I primary total hip replacements implanted in 72 patients between January 1993 and December 1998. The mean follow-up was 8.2 (range 7–12) years. There was significant polyethylene wear and osteolysis associated with the acetabular cup .The cumulative survival of the cup with revision being the end point at 8.2 years was 95.1% (95% CI: 92–97.6%). However, the cumulative survival of the cup with revision and aseptic loosening together was 72% (95% CI: 61–78%) and survival of the acetabular liner for wear was 62% (95% CI: 48–74%). Stem survival with revision being the end point was 100%. In spite of significant radiological failures of the cups, most patients remained asymptomatic. Though results of the ABG stems in this series were good, we advocate a regular follow-up of all these hips in view of the poor outcome of the cups.