The innovative method of high accuracy interferometric calibration of the Precision Laser Inclinometer
2017
In this report a high accuracy method for an interferometric calibration of the Precision Laser Inclinometer (PLI) is proposed. The method is based on the simultaneous measurement of: (a) the PLI base calibration slope (angle), set by a piezoelectric positioner, via laser interferometry and (b) the PLI response signal. A calibration coefficient of 322.5 ± 1.9 μrad/V has been determined experimentally in an interval [4 × 10–7, 4 × 10–6] rad in which there is a linear dependence between the PLI-signal and the calibration angle.