Real-time electrochemical imaging using an individually addressable multi-channel electrode
2000
We developed a real-time electrochemical imaging method that uses a multiple enzyme-modified microelectrode. The method will enable the investigation of the functions of biological materials and cells. To test its effectiveness, we imaged the two-dimensional concentration distribution for hydrogen peroxide and L-glutamate in a standard solution. The multiple electrode consists of an 88 array of 3030 mm 2 carbon micro electrode. Each electrode was connected to a 64-channel potentiostat that could apply a potential to all electrodes at the same time. The multiple electrode was coated with an Os-polyvinylpyridine based polymer (Os-gel) containing horse radish peroxidase (HRP) to detect hydrogen peroxide, which is a very common product of oxidase enzyme. When measuring glutamate, which is a well-known neurotransmitter in the mammalian central nerve system, we modified the electrode with a bilayer of Os-gel-HRP and GluOx. The detection limit of our method was 1 mM and images of the glutamate concentration-distribution changes induced by local injection of glutamate through microcapillary were obtained in real time. © 2000 Elsevier Science S.A. All rights reserved.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
22
References
24
Citations
NaN
KQI