Magnetic resonance imaging-guided balloon angioplasty of coarctation of the aorta: a pilot study

2006 
Background— MRI guidance of percutaneous transluminal balloon angioplasty (PTA) of aortic coarctation (CoA) would be desirable for continuous visualization of anatomy and to eliminate x-ray exposure. The aim of this study was (1) to determine the suitability of MRI-controlled PTA using the iron oxide–based contrast medium Resovist (ferucarbotran) for catheter visualization and (2) to subsequently apply this technique in a pilot study with patients with CoA. Methods and Results— The MRI contrast-to-noise ratio and artifact behavior of Resovist-treated balloon catheters was optimized in in vitro and animal experiments (pigs). In 5 patients, anatomy of the CoA was evaluated before and after intervention with high-resolution respiratory-navigated 3D MRI and multiphase cine MRI. Position monitoring of Resovist-treated catheters was realized with interactive real-time MRI. Aortic pressures were continuously recorded. Conventional catheterization was performed before and after MRI to confirm interventional succe...