Effect of blood contamination on amniotic fluid detection in vitro using immunoassays

2019 
AbstractObjective: To determine the accuracy of Actim PROM®, Amnisure®, and ROM Plus® tests for detecting amniotic fluid proteins in the setting of blood contamination.Methods: IGFBP-1 and AFP are proteins present in high concentrations in amniotic fluid, and are detected by three commercially-available immunoassays used for diagnosing ruptured membranes: Actim PROM®, Amnisure®, and ROM Plus®. We used whole blood samples and diluted these with amniotic fluid (containing known concentrations of amniotic fluid proteins) to whole blood levels of 50, 20, 10, 5, and 1%. Actim PROM®, Amnisure®, and ROM Plus® tests were performed on each sample in duplicate according to package insert instructions. Results were interpreted independently at 5, 10, 15, and 20 min by two obstetricians who were blinded to the concentrations of blood and amniotic fluid proteins in each sample. Results of each test were determined to be true positive, false negative, false positive, or true negative based on physician interpretation a...