Reliability of Compartmental Body Composition Measures in Weight-Stable Adults Using GE iDXA: Implications for Research and Practice

2018 
The aim of this study was to explore the reliability and precision of body compartment measures, in particular visceral adipose tissue, in weight stable adults over a range of BMIs using GE-Lunar iDXA. Weight-stable participants aged 18–65 years had a total body composition scan on GE-Lunar iDXA either on three separate occasions over a three month period (n = 51), or on a single occasion for duplicate scans with repositioning (n = 30). The coefficient of variation (CV%) and least significant change (LSC) of body compartments were calculated. The CV was higher for all measures over three months (range 0.8–5.9%) compared with same-day precision-scans (all 500 g. Changes observed in VAT mass levels below 500 g should be interpreted with caution due to lack of precision and reliability. All other compartmental measures demonstrated good reliability, with less than 6% variation over three months.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    46
    References
    10
    Citations
    NaN
    KQI
    []