Graft flow assessment and early coronary artery bypass graft failure: a computed tomography analysis

2021 
Objectives We evaluated graft patency by computed tomography and explored the determinants of intraoperative mean graft flow (MGF) and its contribution to predict early graft occlusion. Methods One hundred and forty-eight patients under a single surgeon were prospectively enrolled. Arterial and endoscopically harvested venous conduits were used. Intraoperative graft characteristics and flows were collected. Graft patency was blindly evaluated by a follow-up computed tomography at 11.4 weeks (median). Results Graft occlusion rate was 5.2% (n = 22 of 422; 8% venous and 3% arterial). Thirteen were performed on non-significant proximal stenosis while 9 on occluded or >70% stenosed arteries. Arterial and venous graft MGF were lower in females (Parterial = 0.010, Pvenous = 0.009), with median differences of 10 and 13.5 ml/min, respectively. Arterial and venous MGF were associated positively with target vessel diameter ≥1.75 mm (Parterial = 0.025; Pvenous = 0.002) and negatively with pulsatility index (Parterial Conclusions We demonstrate that MGF absolute values are influenced by coronary size, gender and graft type. Intraoperative MGF of >26.5 ml/min for arterial and >36.5 ml/min for venous grafts is the most reliable independent predictor of early graft patency. Modern-era coronary artery bypass graft is associated with low early graft failure rates when transit time flow measurement is used to provide effective intraoperative quality assurance.