On line continuous monitoring of blood lactate in men by a wearable device based upon an enzymatic amperometric lactate sensor

1993 
Abstract A wearable device for the continuous measurement of lactate in the blood was constructed by the combination of continuous blood sampling employing a double lumen catheter with an amperometric lactate sensor. In vitro , the lactate sensor turned out to have a linear concentration range between 0 and 15 mmol/l. The response time of the sensor itself amounted to 100 sec, whereas the lag time for blood sampling amounted to 2·2 min. In vivo , the lactate sensor was successfully used for the detection of changes of the blood lactate concentration following strenuous exercise in 7 healthy volunteers, in two cases up to 22 h. In conclusion, the technique of continuous blood sampling by the use of a double lumen catheter in combination with the amperometric lactate sensor is feasible and simplifies frequent blood lactate estimations.