Permanent His Bundle Pacing in Patients With Congenital Complete Heart Block: A Multicenter Experience.

2020 
Abstract Objectives This study retrospectively assessed the safety and efficacy of permanent His bundle pacing (HBP) in patients with congenital complete heart block (CCHB). Background HBP has become an accepted form of pacing in adults. Its role in CCHB is not known. Methods Seventeen patients with CCHB who underwent successful HBP were analyzed at 6 academic centers between 2016 and 2019. Nine patients had de novo implants, and 8 patients had previous right ventricular (RV) leads. Three RV paced patients had reduced left ventricular ejection fractions at the time of HBP. Implant/follow-up device parameters, New York Heart Association functional class, QRS duration, and left ventricular ejection fraction data were analyzed. Results Patients’ mean age was 27.4 ± 11.3 years, 59% were women, and mean follow-up was 385 ± 279 days. The following parameters were found to be statistically significant between implant and follow-up, respectively: impedance, 602 ± 173 Ω versus 460 ± 80 Ω (p  Conclusions HBP seems to be safe and effective, with improvement in clinical outcomes in patients with CCHB. Larger studies with longer follow-up periods are required to confirm our findings.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    36
    References
    2
    Citations
    NaN
    KQI
    []