Metrological traceability of values for α-amylase catalytic concentration assigned to a commutable calibrator materials

2010 
Abstract Background The International Standard ISO 18153 describes how to assure the metrological traceability of catalytic concentration values assigned to commercial calibration materials following the reference measurement system. We applied this approach to the standardization of α-amylase measurements. Methods Traceable values of catalytic activity with measurement uncertainty were assigned to three commercial calibrator materials using α-amylase primary reference measurement procedure described by the International Federation of Clinical Chemistry and Laboratory Medicine (IFCC). The metrological traceable to the SI units calibration was validated by measuring human serum and plasma samples using the primary reference procedure and two routine commercial measurement procedures using different substrates (4,6-ethylidene(G1)-4-nitrophenyl(G7)-α-(1→4)- d -maltoheptaoside (EPS) and 2-chloro-4-nitrophenyl-α- d -maltotrioside (CNP)) and calibrated with the traceable materials. Previously the commutability of the commercial calibration materials was verified. Results Procedures comparison showed that the primary reference procedure and routine procedures have the same analytical specificity. The commercial calibrators tested showed to be commutable and were used to recalibrate the routine measurement procedures. The agreement of procedures improves after recalibration with commutable calibrator materials. Conclusions The implementation of a reference system for α-amylase measurements was demonstrated that assures the traceability of patients' results to SI units and to harmonize the results obtained by two different routine procedures.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    27
    References
    9
    Citations
    NaN
    KQI
    []