Effect of Algorithm Aggressiveness on the Performance of the Hypoglycemia-Hyperglycemia Minimizer (HHM) System
2014
Background: The Hypoglycemia-Hyperglycemia Minimizer (HHM) System aims to mitigate glucose excursions by preemptively modulating insulin delivery based on continuous glucose monitor (CGM) measurements. The “aggressiveness factor” is a key parameter in the HHM System algorithm, affecting how readily the system adjusts insulin infusion in response to changing CGM levels. Methods: Twenty adults with type 1 diabetes were studied in closed-loop in a clinical research center for approximately 26 hours. This analysis focused on the effect of the aggressiveness factor on the insulin dosing characteristics of the algorithm and, to a lesser extent, on the glucose control results observed. Results: As the aggressiveness factor increased from conservative to medium to aggressive: the maximum observed insulin dose delivered by the algorithm—which is designed to give doses that are corrective in nature every 5 minutes—increased (1.00 vs 1.15 vs 2.20 U, respectively); tendency to adhere to the subject’s nominal basal dose decreased (61.9% vs 56.6% vs 53.4%); and readiness to decrease insulin below basal also increased (18.4% vs 19.4% vs 25.2%). Glucose analyses by both CGM and Yellow Springs Instruments (YSI) indicated that the aggressive setting of the algorithm resulted in the least time spent at levels >180 mg/dL, and the most time spent between 70-180 mg/dL. There was no severe hyperglycemia, diabetic ketoacidosis, or severe hypoglycemia for any of the aggressiveness values investigated. Conclusions: These analyses underscore the importance of investigating the sensitivity of the HHM System to its key parameters, such as the aggressiveness factor, to guide future development decisions.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
9
References
6
Citations
NaN
KQI