Three-dimensional conformal pancreas treatment: comparison of four- to six-field techniques.

1995 
Purpose: We compare practical conformal treatment approaches to pancreatic cancer using 6 and 18 MV photons and contrast those approaches against standard techniques. Methods and Materials: A four-field conformal technique for treating pancreas cancer has been developed using nonopposed 18 MV photons. This approach has been extended to 6 MV photon application by the addition of one to two fields. These techniques have been optimized to increase sparing of normal liver and bowel, compared with opposed-field methods, to improve patient tolerance of high doses. In this study we compare these techniques in a simulated tumor model in a cylindrical phantom. Dose-volume analysis is used to quantify differences between the conformal, nonopposed techniques with conformal, opposed field methods. This model is also used to evaluate the effect of 1-2 cm setup errors on dose-volume coverage. Results: Dose-volume analysis demonstrates that five-to-six field conformal treatments using 6 MV photons provides similar or better dose coverage and normal tissue sparing characteristics as an optimized 18 MV, four-field approach when 1-2 cm margins are included for setup uncertainty. All approaches using nonopposed beam geometry provide significant reduction in the volume of tissue encompassed by the 30-50% isodose surfaces, as compared with four-field box techniques. Conclusions: Three-dimensional (3D) conformal treatments can be designed that significantly improve dose-volume characteristics over conventional treatment designs without costing unacceptable amounts of machine time. Further, deep intraabdominal sites can be adequately accessed and treated on intermediate energy machines with a relatively moderate increase in machine time
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    11
    References
    27
    Citations
    NaN
    KQI
    []