Integration and calibration of DEPFET macropixel detectors for MIXS
2012
The Mercury Imaging X-ray Spectrometer (MIXS) is an instrument on board of the 5th ESA cornerstone mission
BepiColombo. This Spectrometer comprises two instruments for imaging x-ray spectroscopy of the Mercury
surface. The detector plane arrays (DPA) for the energy and spatial resolved detection of x-rays are based on
DEPFET (Depleted P-channel FET) macropixel detectors with 64×64 pixel each and 300×300 μm 2 pixel
size. The MIXS target energy band is from 0.5 to 7 keV with an energy resolution better than 200 eV at 1 keV
at mission end. This allows to access the Fe-L line at about 0.7 keV, which was not accessible to previous
instruments, and to separate the x-ray lines of the elements of interest.
Before a detector chip is integrated into a detector module, it is electrically pre-characterized in order to select
only the best chips for the complex and time-consuming integration. The high degree of complexity of the
integration process comes from the need to thermally decouple the detector chip from its readout and steering
ASICs by a sophisticated mechanical structure, due to the limited amount of cooling power available for the
instrument. After the spectroscopic characterization of the detector modules, the flight and flight spare detectors
were calibrated at the PTB (Physikalisch-Technische Bundesanstalt) beamlines at the BESSY-II synchrotron.
We report on the pre-characterization, integration, qualification and calibration of MIXS flight and flight spare
detectors, which is now successfully completed.