Chromatographic purification of antimony for precise isotope analysis by MC-ICP-MS

2020 
A two-step method for purifying Antimony (Sb) from geological samples for precise measurement of Sb isotopic composition using multi-collector inductively coupled plasma mass spectrometry (MC-ICP-MS) was developed. Sb was separated from matrix using two successive chromatographic columns loaded with AG 1-X4 and AG 50W-X8 resin, respectively. The separation method was calibrated using geological samples by assessing the extraction yield and the reproducibility of the measured isotopic composition, and resulted in a relatively lower procedure blank and better recovery (98.7% ± 5.6%, n=8, 2SD). The validation of the whole protocol, together with instrumental analysis, was confirmed by the investigation of the matrix effect and the result of a standard addition experiment. A combined sample-standard bracketing and external normalization (SSB-EN) method using Cd as an internal element was used for mass bias correction, which gives a reproducibility better than 0.04‰ (2SD). A variation of 0.29‰ of δ123Sb value (relative to NIST SRM 3102a standard) was observed in artificial and geological samples, implying the potential application of Sb isotopes in various environments.