Abstract This paper describes a compact cooling system using a single-stage stirling-type cryocooler for a practical HTS sampler. The system was designed to cool down an HTS sampler module below 50 K, enabling a bandwidth of the chip more than 100 GHz. The system measures 150 mm in width, 140 mm in height and 310 mm in depth, and weighs 5 kg. Semi-rigid coaxial cables made of brass with a silver coated inner conductor were adopted for a signal to be measured and a trigger pulse. The loss for the signal line was less than 1.5 dB at 50 GHz with relatively small thermal inflow. Thermal inflows from low frequency lines, IF signal lines for control/output of the sampler and dc bias lines, were minimized by choosing proper wires. A new sampler module with reduced weight was placed on the cold stage, which was surrounded by double magnetic shields. The module was successfully cooled down to less than 50 K with cooling time of 1 h in the system. We have also succeeded in observing sinusoidal waveforms with the HTS sampler cooled by the compact cooling system.