Time resolved laser induced fluorescence measurements: Considerations when using Nd:YAG based system

2012 
Abstract Time-resolved laser-induced fluorescence (TR-LIF) and the laser induced breakdown spectroscopy (LIBS) have been shown to be methods which are fast and sensitive to provide information about the constituents in analyzed samples. TR-LIF and LIBS have similar hardware requirements. In this paper, we analyze some characteristics of TR-LIF/LIBS system implemented in our laboratory, considering the fact that the excitation part of the system is based on Nd:YAG laser and Optical Parametric Oscillator (OPO). The laser is more than powerful enough (365 mJ at 1064 nm, variable OPO output >5 mJ) for LIBS, but somehow slow (the length of fundamental laser harmonic output pulse is about 5 ns) for fluorescence measurements in our present area of interest, namely plants and food products. Fortunately, the pulse length of tunable OPO output (320–475 nm) is less then 1 ns, so by means of a correct deconvolution procedure it is possible to measure the fluorescence lifetimes in the range as small as a few nanoseconds. The fluorescence detection part of our system is based on picosecond streak camera. Using the fluorescent dyes (Rhodamine B and Fluorescein) ethanol solutions we verified the analyzing capabilities of our TR-LIF system.