A multi-reader in vitro study using porcine kidneys to determine the impact of integrated circuit detectors and iterative reconstruction on the detection accuracy, size measurement, and radiation dose for small (<4 mm) renal stones

2017 
BackgroundDetection of small renal calculi has benefitted from recent advances in computed tomography (CT) scanner design. Information regarding observer performance when using state-of-the-art CT scanners for this application is needed.PurposeTo assess observer performance and the impact of radiation dose for detection and size measurement of <4 mm renal stones using CT with integrated circuit detectors and iterative reconstruction.Material and MethodsTwenty-nine <4 mm calcium oxalate stones were randomly placed in 20 porcine kidneys in an anthropomorphic phantom. Four radiologists used a workstation to record each calculus detection and size. JAFROC Figure of Merit (FOM), sensitivity, false positive detections, and calculus size were calculated.ResultsMean calculus size was 2.2 ± 0.7 mm. The CTDIvol values corresponding to the automatic exposure control settings of 160, 80, 40, 25, and 10 Quality Reference mAs (QRM) were 15.2, 7.9, 4.2, 2.7, and 1.3 mGy, respectively. JAFROC FOM was ≥ 0.97 at ≥ 80 QRM, ...