Mode of recording and modulation frequency effects of auditory steady state response thresholds

2017 
Abstract Introduction The performance of auditory steady state response (ASSR) in threshold testing when recorded ipsilaterally and contralaterally, as well as at low and high modulation frequencies (MFs), has not been systematically studied. Objective To verify the influences of mode of recording (ipsilateral vs . contralateral) and modulation frequency (40 Hz vs . 90 Hz) on ASSR thresholds. Methods Fifteen female and 14 male subjects (aged 18–30 years) with normal hearing bilaterally were studied. Narrow-band CE-chirp ® stimuli (centerd at 500, 1000, 2000, and 4000 Hz) modulated at 40 and 90 Hz MFs were presented to the participants’ right ear. The ASSR thresholds were then recorded at each test frequency in both ipsilateral and contralateral channels. Results Due to pronounced interaction effects between mode of recording and MF ( p Conclusions Based on the findings, the present study recommends two different protocols for an optimum threshold testing with ASSR, at least when testing young adults. This includes the use of contra-40 Hz recording mode due to its promising performance in hearing threshold estimation.
    • Correction
    • Source
    • Cite
    • Save
    • Machine Reading By IdeaReader
    24
    References
    2
    Citations
    NaN
    KQI
    []