Evaluation of Microwave Ablation Efficacy by Strain Elastography and Shear Wave Elastography in ex Vivo Porcine Liver.

2021 
Abstract The aim of this study was to evaluate the efficacy of microwave ablation by ultrasound (US), strain elastography (SE) and shear-wave elastography (SWE). An ex vivo model of porcine liver was adopted. According to ablation power and duration, 30 samples were divided into three groups: group 1 (45 W, 30 s), group 2 (45 W, 15 s) and group 3 (30 W, 30 s). US was used to measure the largest transverse diameter (D1), vertical diameter (D2) and anteroposterior diameter (D3) of the ablated area. SE was used to measure the largest transverse diameter (SEL1), vertical diameter (SEL2) and anteroposterior diameter (SEL3). The actual size of the ablated area was measured as the largest transverse diameter (L1), vertical diameter (L2) and anteroposterior diameter (L3). SWE values and temperatures were measured in the central lesion (region a), marginal area (region b) and unablated area (region c). At 1 h post-ablation, the values measured by US (D1, D2, D3) were all significantly smaller than the ablated area (L1, L2, L3) in all three groups. Except for SEL2 in group 1, there was no significant difference in the results between SEL and L among the three groups. All SWE results were significantly higher post-ablation than pre-ablation in the central lesion (region a) and marginal area (region b, all p values