A comparison of the antegrade core trap and reverse bevel needles for EUS-guided fine-needle biopsy sampling of liver mass: a prospective randomized cross over study.

2021 
BACKGROUND Data on the use of EUS-guided fine-needle biopsy (EUS-FNB) of solid liver mass (SLM) for pathology is limited. METHODS To prove superiority of the diagnostic rate of the newly designed modified Menghini-type needle with a beveled side-slot near the needle tip with slot cutting edge directed 20-gauge antegrade bevel (group A) over the original 22-gauge reverse bevel (group B) for EUS-guided fine-needle biopsy (EUS-FNB) of solid liver mass (SLM) in a prospective crossover randomized controlled trial. RESULTS The overall diagnostic accuracy rate of the 52 passes was 86.5% (45/52) and of group A versus B were 88.5% (23/26) versus 84.6% (22/26), respectively, p = 0.858. Tissue adequacy levels of both groups were not significantly different (grade A: B: C = 18:6:2 versus 16:7:3), p = 0.839). Grading of blood contamination of both groups was not significantly different. However, it was found that the group-A needles could biopsy tissue of significantly longer length than that of the group B; 1.3 cm (SD = 0.76) versus 0.8 cm (SD = 0.54); p = 0.007. CONCLUSION The use of EUS-FNB of SLM is highly effective with similar levels of efficacy and number of adverse events between both types of needles. THE TRIAL REGISTRATION NUMBER Thai Clinical Trial Registration No. TCTR2018081002.