A micro-SQUID with dispersive readout for magnetic scanning microscopy

2018 
We have designed and characterized a micro-SQUID with dispersive readout for use in low temperature scanning probe microscopy systems. The design features a capacitively shunted RF SQUID with a tunable resonance frequency from 5 to 12 GHz, micrometer spatial resolution, and integrated superconducting coils for local application of magnetic fields. The SQUID is operated as a nonlinear oscillator with a flux- and power-dependent resonance frequency. Measurements for device characterization and noise benchmarking were carried out at 4 K. The measured flux noise above 10 kHz at 4 K is 80 nΦ0 Hz−1∕2 at a bandwidth of 200 MHz. Estimations suggest that one can benefit from parametric gain based on inherent nonlinearity of the Josephson junction and reduce the flux noise to 30 nΦ0Hz–1∕2 at 100 mK, which corresponds to 10.6 μBHz–1∕2 for a magnetic moment located at the center of the pickup loop.