Cryogenic quartz crystal microbalance
1975
A radiatively cooled Cryogenic Quartz Crystal Microbalance designed to monitor highly volatile contaminants on the shuttle is described. Measurements are made with two 15-MHz microbalances having removable, optically polished sensors mounted in a radiant cooler. One sensor operates below the freezing point of water and monitors contamination including that of water vapor. The second sensor is heated and monitors the contamination background. It provides a reference from which the density of the water vapor cloud enveloping the shuttle is determined. The design incorporates a low-power dissipation oscillator, heaters for ice removal, and a method for attaching second-surface mirrors to the radiator employing an indium type solder instead of a room temperature vulcanizer.
Keywords:
- Correction
- Source
- Cite
- Save
- Machine Reading By IdeaReader
0
References
0
Citations
NaN
KQI